Location -> Products -> Analyzer Instrument -> On-line Instrument -> Industry Conductivity Meter

Type:CX-ICM-1101A industrial conductivity meter
Illustrate:It is applicable for pure and ultrapure water treatment, industrial and sewage water treatment in the following industries: electric power, environmental protection, chemical industry, petroleum, paper making, metallurgy, semi-conductor, medicine, foodstuff, etc
CX-ICM-1101A industrial conductivity meter
Brief Introduction1.It is a kind of on-line monitoring instrument for water quality analysis equipped with a microprocessor.
2.Widely used for continuous monitoring of the conductivities of demineralized water, raw water, steam condensate, distilled sea water and de-ionized water.
Main Features
1.Large vacuum fluorescent display (VFD), and man-machine interaction function.
2.The functions of automatic temperature compensation and automatic range-switching.
3.Available with history curve, real-time trend chart and clock.
4.There are four kinds of built-in recorders. It has the functions of isolated current output and alarm.
5.The user can set the password or consult the “Help” menu or the “Service Guide” Menu.
6.It has power-off protection. The data can be saved for 5 years under a continuous power-failure condition.
7.The communication interface can be connected directly to a computer and enables remote operation and control.
8.The instrument adopts universal module power supply.
9.It is equipped with rib-type, plug-in and push-in types of conductivity electrode with a electrode constant of
K=0.01, K=0.1 or K=1.0.
10.It adopts titanium electrode that can withstand oxidization and erosion and can be easily used and maintained.
Main Parameters
Measuring range | 0~ 20 μS.cm-1 ,0~200 μS.cm-1, 0~2000μS.cm-1, 0~20mS.cm-1 |
Automatic temperature compensation | 0-60℃ (with 25℃ as the reference temperature) |
Intrinsic error | Conductivity:±0.5%F.S±1 unit |
Error of the temperature compensation | ±0.5%F.S |
Stability | ±0.2/F.S±1 unit/24h |
Repeatability error | ≤0.2%F.S±1 unit |
Output current error | ±1%F.S |
Alarm error of the electronic units | ±2%F.S |
Water sample pressure | 0.6MPa |
Intrinsic error of the instrument: conductivity | ±1%F.S, temperature: ±0.5℃ |
Maximum output load and error |
For 0~10mA, the maximum output load is l.5K Ω±1%F.S. For 4~20mA, the maximum output load is 750Ω±1%F.S. |
Power supply | 85~265VAC 50±1Hz |
Overall dimension | 151 (length) x 151 (width) x 251 (height) mm |
Dimension of the mounting hole | 138 x 138mm |
Connection pattern of inlet/ outlet water pipes | Φ6,Φ8,Φl0,Φl2 and Φ14. |
Weight | 1.2kg |
sensor types and installation methods | Flow-through type,duct-type,immersed type |
Last:Without the Next:CX-ICM-1101 industrial conducti..