Type:CX-ICM-1107 industrial conductivity meter
Illustrate:It is widely used for on-line continuous monitoring of pure water in pharmacy, electronics, chemical engineering and many other industries, and is an optimal supporting instrument for various medium and small-sized pure water equipment.
CX-ICM-1107 Industrial Conductivity Meter
Brief Introduction
1.It is an on-line measurement and control instrument for water quality in industrial processes.
2.It integrates measurement of conductivity, resistivity and temperature.
3.Under the single-chip technology, it provides discretionary setting of temperature factor of pure water, constant of electrode, upper alarming threshold of conductivity and lower alarming threshold of resistivity by means of keystroke.
4.This meter can be used as a supporting instrument for the front conductance and back resistance between the secondary reverse osmosis and the EDl ion exchange resin mixed-bed.
Main Features
1.Four-level measuring range, automatic switching, automatic carry-bit display, simple operation and highmeasuring precision.
2.Automatic temperature measurement and automatic temperature compensation for solution.
3.3-digit LCD digital display, simultaneous display of temperature and measured value, beautiful outline and clear display.
4.Solution temperature factor and electrode constant can be set discretionarily.
5.Alarming upon and control over excess conductivity and resistivity (conductivity 0-1999 μS·cm-1)
6.Resistivity (0-18.25 MΩ·Cm) Action points of relay can be set within the measuring range.
7.Isolated 4-20mA current signal output is provided for resistivity (Channel B), which can be used as the output of the meter or realized through transmitter.
Main parameters
Measuring range |
Channel A conductivity: 0-2000μS·cm-1 with automatic range change Channel B resistivity: 0~18.25MΩ·cm with automatic range change Temperature measuring range: 0-60°C |
Intrinsic error of electronic units | ±1.0%(FS) ±1 unit |
Intrinsic error of the instrument: | ±2%(FS) ±1 unit |
Temperature compensation range | 5-50 °C, with a base temperature of 25°C |
Temperature measurement error | ±1°C |
Electrode |
For Channel A conductivity measurement: l.0 cm-1; material: stainless steel For Channel B resistivity measurement: 0.01cm-1; material: stainless steel |
Working pressure | 0-0.5Mpa |
Output current signal (Channel B) | Isolated output of 4-20mA (with a max. load of 500Ω) |
Capacity of alarming contact | 2A/250V AC (on resistance load) |
Size of thread: | 1/2” pipe thread |
Cable length | 5m or as specified |
Overall dimension |
96 x 96 x 130mm(length x width x depth); dimension of the hole: 92 x 92mm |
Working conditions | |
Ambient temperature | 5-50 °C |
Relative humidity | No more than 85% |
Power supply | AC220V±10% 50Hz |
No remarkable vibration | |
No magnetic field interference other than terrestrial magnetic field |
Last:CX-ICM-1105A industrial conduct.. Next:CX-ICM-1105B industrial conduct..