Type:1301 series Industrial Dissolved Oxygen Meter

Illustrate:CX-IDO-1301 Industrial Dissolved Oxygen Meter is a kind of high-intelligence in-line meter that is developed by our company and used for continuous on-line monitoring.

CX-IDO-1301 Industrial Dissolved Oxygen Meter

 

CX-IDO-1301 Industrial Dissolved Oxygen Meter is a kind of high-intelligence in-line meter that is developed by our company and used for continuous on-line monitoring. It can be equipped with CX-IDO-120F Polarographic Electrode and can automatically switch from ppb level to ppm level of wide-range measurement. This instrument is designed for monitoring the oxygen content in the boiler feed water, condensate water and sewage.

Features:  

All the data are displayed in Chinese. It can be easily operated:

It has a complete Chinese display and elegant interface: Liquid crystal display module with high resolution is

adopted. All the data, status and operation prompts are displayed in Chinese. There is no symbol or code that is

defined by the manufacturer.

Simple menu structure and text-type man-instrument interaction: Compared with traditional instruments,

CX-IDO-1301has many new functions. As it adopts classified menu structure, which is similar to that of a

computer, it is clearer and more convenient. It is not necessary to remember the operation procedures and

sequences. It can be operated according to the prompts on the screen without the guidance of an operation manual.

Multi-parameter display: The oxygen concentration value, input current (or output current), temperature values,

time and status can be displayed on the screen at the same time. The main display can show the oxygen

concentration value in a 10 x 10mm size. As the main display is eye-catching, the displayed values can be seen

from a long distance away. Six sub-displays can display such information as the input or output current,

temperature, status, week, year, day, hour, minute and second, so as to adapt to different users’ habits and to

conform with different reference times set by the users.

 

Technical indexes:

1. Measuring range: 0~100.0ug/L; 0~20.00 mg/L (automatic switching); 0~60%

Resolution: 0.1ug/L; 0.01 mg/L; 0.1℃

2. Intrinsic error of the whole instrument: ug/L: ±l.0%FS; mg/L: ±0.5%FS, temperature: ±0.5℃

3. Repeatability of the indication of the whole instrument: ±0.5%FS

4. Stability of the indication of the whole instrument: ±1.0%FS

5. Automatic temperature compensation range: 0~60℃, with 25℃ as the reference temperature.

6. Response time: <60s (98% and 25℃ of the final value) 37℃: 98% of the final value < 20 s

7. Clock accuracy: ±1 minute/month

8. Output current error: ≤±l.0%FS

9. Isolated output: 0-10mA (load resistance <15KΩ); 4-20mA (load resistance <750Ω)

10. Communication interface: RS485 or 232(optional)

11. Data storage capacity: l month (1 point/5 minutes)

12. Saving time of data under a continuous power-failure condition: 10 years

13. Alarm relay: AC 220V, 3A

14. Power supply: 220V±10% 50±1HZ

15. Overall dimension: secondary meter: 146 (length) x 146 (width) x 108 (depth) mm;

dimension of the hole: 138 x 138mm

16. Weight: 1.0kg (secondary meter)

17. Working conditions: ambient temperature: 0-60℃; relative humidity <85%

18. The connection tubes for inlet and outlet water: Pipes and hoses should be of the following three sizes: Φ8,

Φ10 and Φl2.

Last:1303 series Industrial Dissolve.. Next:Industrial Dissolved Oxygen Met..