Location -> Industrial online analyzer -> 4.Conductivity Meter

Type:1101 series Industrial Conductivity Meter
Illustrate:CX-ICM-1101 Industrial Conductivity Meter is our company’s latest top-class instrument equipped with a microprocessor and complete Chinese menu.
CX-ICM-1101 Industrial Conductivity Meter
CX-ICM-1101 Industrial Conductivity Meter is our company’s latest top-class instrument equipped with a microprocessor and complete Chinese menu. It has Chinese display and Chinese menu operation pattern; it is characterized by its intelligent operation, multi-function and high measuring performance and its strong environmental adaptability. It can be widely used for continuous monitoring of the conductivities of solutions used in the following fields: thermal power plant, chemical fertilizer, chemical industry, metallurgy, environmental protection, pharmacy manufacturing, biochemistry, foodstuff and running water.
Features:
It has a complete Chinese display and friendly interface. Various parameters
can be displayed at the same
time: conductivity, output current, temperature, time and status. Bitmap type
liquid crystal display module
with high resolution is adopted. All the data, status and operation prompts
are displayed in Chinese. There
is no symbol or code that is defined by the manufacturer.
Technical indexes:
1.Conductivity measuring range: 0.01~20μS/cm(Electrode: K=0.01)
00.1~200μS/cm (Electrode: K=0.1)
1.0~2000μS/cm (Electrode: K=1.0)
10~20000μS/cm (Electrode: K=10.0)
30~600.0mS/cm (Electrode: K=30.0)
2. Intrinsic error of the electronic unit: conductivity: ±0.5%FS, temperature: ±0.30℃
3. Range of automatic temperature compensation: 0~99.9℃, with 25℃
as the reference temperature
4. Water sample tested: 0~99.9℃, 0.6MPa
5. Intrinsic error of the instrument: conductivity: ±1.0%FS, temperature: ±0.5℃
6. Automatic temperature compensation error of the electronic unit: ±0.5%FS
7. Repeatability error of the electronic unit: ±0.2%FS±1 Unit
8. Stability of the electronic unit: ±0.2%FS±1 unit/24h
9. Isolated current output: 0~10mA ( load<1.5kΩ)
4~20mA (load<750Ω)
10. Output current error: ≤±l%FS
11. Error of electronic unit caused by ambient temperature: ≤±0.5%FS
12. Error of the electronic unit caused by the supply voltage: ≤±0.3%FS
13. Alarm relay: AC 220V, 3A
14. Communication interface: RS485 or 232 (optional)
15. Power supply: AC 220V±22V, 50Hz±1Hz
16. Protection grade: IP65
17. Clock accuracy: ±1 minute/month
18. Data storage capacity: l month (1 point/5 minutes)
19. Saving time of data under a continuous power-failure condition: 10 years
20. Overall dimension: 146 (length) x 146 (width) x 108 (depth) mm; dimension
of the hole: 138 x 138mm
21. Working conditions: ambient temperature: 0~60℃; relative humidity <85%
22. Weight: 1.0kg
23. The conductivity electrodes with the following five constants are usable:
K=0.01, 0.1, 1.0, 10.0, and 30.0.
Last:1103 seriesIndustrial Conducti.. Next:1101A series Industrial Conduc..